Best Practices for Ultra-Low Temperature Freezers

Ultra-Low Temperature Freezers (ULTs) are essential equipment for researchers and scientists. These ULTs require a significant amount of energy to maintain low temperatures. Improper care of ULTs can lead to elevated energy costs, damage to samples, and leakage of greenhouse gases. Below is a guide of best practices in order to lighten your ULT’s energy burden and maximize its life.

Filters

  • Filters should be inspected monthly. 
  • Clean the filter by vacuuming the filter door to remove dust. Wash the filter with soap and water. Let it air dry before reinserting.

Gasket

  • Inspect the door gasket for tears and deformities regularly.
  • Wipe frost from around the gasket and between the folds.

Defrosting

ULTs should be defrosted regularly.

  1. Remove samples.
  2. Shut down unit.
  3. Prop open doors & melt all ice.
  4. Completely wipe & dry unit. 
  5. Reach set point temperature & replace samples.

Samples

  • Keep samples organized.
  • Use an inventory management system.
  • Clean samples out regularly.

 Filling

  • Pre-cool samples in a -20ºC freezer before placing them into a ULT.
  • Don’t overfill.

Air Flow 

  • Freezers should be tightly packed together  
  • Keep 8” of clear space above the freezer. 
  • Keep 5” of space behind.  

Room 

  • Place freezers near the room’s return vent.  
  • Rooms at cooler temperatures (<72ºF) can ease the energy load on the ULT. 
  • Don’t place ULTs in hallways.

Operations

  • Consider increasing the set point to -70ºC.
  • Research on the benefits of raising the temperature:“-70 is the New -80”

Frost

  • Regularly wipe frost off of surfaces with a soft cloth.
  • A brush or scraper can be used to clean harder to reach places or more dense frost.
  • Be careful to not puncture the door gasket.

Resources

A presentation from ThermoFisher on ULT best practices which also includes a 12-point maintenance checklist.

Eppendorf has an excellent 1-page maintenance guide.

The National Institutes of Health did a case study on the factors that affect the performance, energy consumption, and carbon footprint of ULTs.

Store Smart’s: Everything You Wanted to Know about Running an Ultra Low Temperature Freezer Efficiently but Were Afraid to Ask

ThermoFisher created a guide to maintenance and upgrades.

My Green Lab has an excellent general resource page. You can find resources on preventative maintenance, increasing set-point temperature, and energy efficiency.

My Green Labs has created a comprehensive list of resources that can assist you in evaluating sustainability metrics such as tax rebates, energy efficiency, carbon footprint, and more when purchasing a new ULT.